Endurant — Class II medical device recall Z-0657-2022

Ongoing recall by Medtronic Vascular, Inc., reported 2022-03-02, distributed nationwide. Stent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip t

Recall numberZ-0657-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic Vascular, Inc., Santa Rosa, CA, United States
Recall initiated2021-12-21
Classified2022-02-18
Reported by FDA2022-03-02
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KE, KR, KW, KY, LB, LT, LU, LV, MA, MT, MX, MY, NL, NO, NZ, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UA, UY, VN, ZA
Quantity14,629 devices
Reason classesdevice malfunction
UPC / GTIN / UDI-DI00613994990969, 00613994990976, 00613994990983, 00613994991027, 00613994991034, 00613994991041, 00613994991133, 00613994991164, 00613994991195, 00613994991201, 00613994991218, 00613994991225, 00613994991249, 00613994991263, 00613994991270, 00613994991294, 00613994991300, 00613994991324, 00613994991423, 00613994991430, 00613994991447, 00613994991454, 00613994991461, 00613994991478, 00613994991485 and 135 more
Serial numbers00613994990969, 00613994990976, 00613994990983, 00613994991027, 00613994991034, 00613994991041, 00613994991133, 00613994991164, 00613994991195, 00613994991201, 00613994991218, 00613994991225, 00613994991249, 00613994991263, 00613994991270, 00613994991294, 00613994991300, 00613994991324, 00613994991423, 00613994991430, 00613994991447, 00613994991454, 00613994991461, 00613994991478, 00613994991485 and 475 more
Model numbersENBF2313C120EE, ESBF2314C103E, ETBF2313C124E

Product

Endurant, Endurant II and Endurant IIs Stent Graft System: Medtronic Endurant Stent Graft System Model/REF ENBF2313C120EE; ENBF2313C120EE; ENBF2313C145EE; ENBF2313C170EE; ENBF2316C120EE; ENBF2316C145EE; ENBF2316C170EE; ENBF2513C120EE; ENBF2513C145EE; ENBF2513C170EE; ENBF2516C120EE; ENBF2516C145EE; ENBF2516C170EE; ENBF2813C145EE; ENBF2813C170EE; ENBF2816C120EE; ENBF2816C145EE; ENBF2816C170EE; ENBF3216C120EE; ENBF3216C145EE; ENBF3216C170EE; ENBF3616C145EE; ENBF3616C170EE; ENCF2323C45EE; ENCF2525C45EE; ENCF2828C45EE; ENCF3232C45EE; ENCF3636C45EE; ENTF2323C70EE; ENTF2525C70EE; ENTF2828C70EE; Medtronic Endurant II Stent Graft System Model/REF ETBF2313C124E; ETBF2313C124EE; ETBF2313C145EE; ETBF2313C145EJ; ETBF2313C166EE; ETBF2313C166EJ; ETBF2316C124E; ETBF2316C124EE; ETBF2316C124EJ; ETBF2316C145E; ETBF2316C145EE; ETBF2316C145EJ; ETBF2316C166E; ETBF2316C166EE; ETBF2316C166EJ; ETBF2513C124E; ETBF2513C124EE; ETBF2513C124EJ; ETBF2513C145EE; ETBF2513C145EJ; ETBF2513C166E; ETBF2513C166EE; ETBF2513C166EJ; ETBF2516C124EE; ETBF2516C145E; ETBF2516C145EE; ETBF2516C145EJ; ETBF2516C166E; ETBF2516C166EE; ETBF2516C166EJ; ETBF2813C124EE; ETBF2813C124EJ; ETBF2813C145E; ETBF2813C145EE; ETBF2813C145EJ; ETBF2813C166E; ETBF2813C166EE; ETBF2813C166EJ; ETBF2816C124EE; ETBF2816C124EJ; ETBF2816C145E; ETBF2816C145EE; ETBF2816C145EJ; ETBF2816C166E; ETBF2816C166EE; ETBF2816C166EJ; ETBF2820C124EE; ETBF2820C145EE; ETBF2820C145EJ; ETBF2820C166E; ETBF2820C166EE; ETBF2820C166EJ; ETBF3216C124E; ETBF3216C124EE; ETBF3216C124EJ; ETBF3216C145E; ETBF3216C145EE; ETBF3216C145EJ; ETBF3216C166EE; ETBF3216C166EJ; ETBF3220C124EE; ETBF3220C145EE; ETBF3220C166E; ETBF3220C166EE; ETBF3220C166EJ; ETBF3616C145E; ETBF3616C145EE; ETBF3616C145EJ; ETBF3616C166E; ETBF3616C166EE; ETBF3616C166EJ; ETBF3620C145EE; ETBF3620C166EE; ETBF3620C166EJ; ETCF2323C49E; ETCF2323C49EE; ETCF2525C49E; ETCF2525C49EE; ETCF2525C49EJ; ETCF2828C49E; ETCF2828C49EE; ETCF2828C49EJ; ETCF3232C49E; ETCF3232C49EE; ETCF3232C49EJ; ETCF3636C49E; ETCF3636C49EE; ETCF3636C49EJ; ETTF2323C70E; ETTF2323C70EE; ETTF2323C70EJ; ETTF2525C70E; ETTF2525C70EE; ETTF2525C70EJ; ETTF2828C70E; ETTF2828C70EE; ETTF2828C70EJ; ETTF3232C70E; ETTF3232C70EE; ETTF3232C70EJ; ETTF3636C70E; ETTF3636C70EE; ETTF3636C70EJ; ETUF2314C102E; ETUF2314C102EE; ETUF2314C102EJ; ETUF2514C102E; ETUF2514C102EE; ETUF2514C102EJ; ETUF2814C102E; ETUF2814C102EE; ETUF2814C102EJ; ETUF3214C102E; ETUF3214C102EE; ETUF3614C102E; ETUF3614C102EE; Medtronic Endurant IIs Stent Graft System Model/REF: ESBF2314C103E; ESBF2314C103EE; ESBF2314C103EJ; ESBF2514C103E; ESBF2514C103EE; ESBF2514C103EJ; ESBF2814C103E; ESBF2814C103EE; ESBF2814C103EJ; ESBF3214C103E; ESBF3214C103EE; ESBF3214C103EJ; ESBF3614C103E; ESBF3614C103EE; ESBF3614C103EJ; Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Reason for recall

Stent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery system. Detachment of the taper tip during the implant procedure may lead to a secondary intervention to attempt to remove the taper tip, either by endovascular retrieval or surgical conversion. There is also potential that the taper tip is intentionally left behind in the patient, at the physician's discretion, if not easily removable.