Data Foundry

Medical device recalls 2022

NobelReplace Conical Connection RP 4 — Class II medical device recall Z-0658-2022

Terminated recall by Nobel Biocare Usa Llc, reported 2022-03-02, distributed nationwide. Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the

Recall numberZ-0658-2022
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmNobel Biocare Usa Llc, Yorba Linda, CA, United States
Recall initiated2021-12-17
Classified2022-02-18
Reported by FDA2022-03-02
Terminated2024-10-25
DistributedNationwide (US)
Countries outside the USCA, MX, TR
Quantity702 units
UPC / GTIN / UDI-DI07332747160529
Lot numbers13116495, 13116496, 13116497
Model numbers36707

Product

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Reason for recall

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0658-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.