Data Foundry

Medical device recalls 2013

Emit — Class II medical device recall Z-0664-2013

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2013-01-16, distributed in CA, FL, IL, KS, MA, MI, MN, MS…. Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R)

Recall numberZ-0664-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2012-12-10
Classified2013-01-10
Reported by FDA2013-01-16
Terminated2014-08-27
DistributedCA, FL, IL, KS, MA, MI, MN, MS, NC, NM, NY, OH, OR, PA, TN, TX, UT, WA
Quantity1,979 cartons
Lot numbers1237, 1264

Product

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL). Homogeneous enzyme immunoassay intended for use in the quantitative analysis of Phenobarbital in human serum or plasma.

Reason for recall

Firm has confirmed higher imprecision with the recalled lots on Beckman Coulter AU(R) Systems and the Syva(R) Emit(R) 2000 Phenobarbital (4D019UL) lots E1, E2, and E3 as an application if the AU(R) System is used, the issue may be experienced. Firm has found that within-run imprecision of the recalled lots frequency of outliers that differ from the expected value by 20%. This is an expansion of