Data Foundry

Medical device recalls 2016

GE Healthcare — Class II medical device recall Z-0679-2016

Terminated recall by GE Medical Systems, LLC, reported 2016-01-27, distributed nationwide. The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which co

Recall numberZ-0679-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Medical Systems, LLC, Waukesha, WI, United States
Recall initiated2015-12-15
Classified2016-01-21
Reported by FDA2016-01-27
Terminated2016-09-21
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BO, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HU, ID, IE, IL, IN, IQ, IS, IT, JO, JP, KE, KR, KW, LK, LT, LV, MA, MX, MY, NI, NL, NO, NZ, OM, PE, PH, PK, PL, QA, RO, RS, SA, SE, SG, SK, TH, TR, TW, VE, VN, ZA
Quantity84

Product

GE Healthcare, Aestiva MRI, 1006-9310-000. This version of the Datex-Ohmeda 7900 ventilator is used in Datex-Ohmeda Aestiva/5 anesthesia systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The Aestiva/5 MRI anesthesia system provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment.

Reason for recall

The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0679-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.