Galaxy bronchoscope — Class II medical device recall Z-0679-2025

Ongoing recall by Noah Medical, reported 2024-12-18, distributed nationwide. Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm

Recall numberZ-0679-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmNoah Medical, San Carlos, CA, United States
Recall initiated2024-10-25
Classified2024-12-12
Reported by FDA2024-12-18
DistributedNationwide (US)
Quantity26
UPC / GTIN / UDI-DI00850048825048
Serial numbersCRB-1502, CRB-1503, CRB-1508, CRB-1513, CRB-1514, CRB-1515, CRB-1516, CRB-1518, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1528, CRB-1530, CRB-1532, CRB-1534, CRB-1535, CRB-1537, CRB-1538, CRB-1541, CRB-1543, CRB1540 and 1 more
Model numbers10000752

Product

Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

Reason for recall

Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.