Data Foundry

Medical device recalls 2024

Alinity c Processing Module REF 03R67-01 The Alinity c Processing… — Class II medical device recall Z-0681-2024

Ongoing recall by Abbott Laboratories, reported 2024-01-17, distributed in AK, AL, AR, AZ, CA, CO, CT, DE…. The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural

Recall numberZ-0681-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbbott Laboratories, Irving, TX, United States
Recall initiated2023-10-19
Classified2024-01-08
Reported by FDA2024-01-17
DistributedAK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
Countries outside the USBR
Quantity1,127 systems
Reason classeslabeling
UPC / GTIN / UDI-DI00380740137380
Model numbers03R67-01

Product

Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c Processing Module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c Processing Module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).

Reason for recall

The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).