Data Foundry

Medical device recalls 2016

Lineage Liners: REF 70102846 — Class III medical device recall Z-0683-2017

Terminated recall by MicroPort Orthopedics Inc., reported 2016-12-07, distributed nationwide. Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal

Recall numberZ-0683-2017
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMicroPort Orthopedics Inc., Arlington, TN, United States
Recall initiated2016-06-21
Classified2016-12-01
Reported by FDA2016-12-07
Terminated2017-07-01
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BE, BG, BR, CA, CH, CN, CY, CZ, DE, DK, ES, FI, FR, GB, GR, IR, IT, JM, JP, KR, LT, MY, NL, PH, PL, PT, RU, SI, TR, TW, VN, ZA
Quantity5,009 units
Serial numbers0101009599, 0101019757, 0101048257, 015140898, 015140899, 015145854, 015155783, 015167730, 016302354, 016303424, 016308023, 016311871, 016311872, 018507062, 018507065, 018507067, 019744517, 019780244, 0201044697, 0201049093, 022A147720, 022A147721, 022A147722, 022A147723, 022A147724 and 379 more
Model numbers70102846, 70202852, 70302858, 70303258, 70402864, 70403264, 70403664

Product

Lineage Liners: REF 70102846, REF 70202852, REF 70202852, REF 70302858, REF 70303258, REF 70402864, REF 70403264, REF 70403664 For use with the conserve metal head

Reason for recall

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0683-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.