DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510… — Class II medical device recall Z-0685-2013

Terminated recall by DePuy Spine, Inc., reported 2013-01-23, distributed nationwide. A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulti

Recall numberZ-0685-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDePuy Spine, Inc., Raynham, MA, United States
Recall initiated2012-07-24
Classified2013-01-16
Reported by FDA2013-01-23
Terminated2013-12-30
DistributedNationwide (US)
Countries outside the USBE, EG, GB, IE, IT, PT
Quantity140 units

Product

DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.

Reason for recall

A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.