Data Foundry

Medical device recalls 2019

VITROS HBeAg Reagent — Class II medical device recall Z-0688-2019

Ongoing recall by ORTHO-CLINICAL DIAGNOSTICS, reported 2019-01-09, distributed nationwide. The firm is extending their previous July 2018 recall to include additional product codes. Biased results may

Recall numberZ-0688-2019
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmORTHO-CLINICAL DIAGNOSTICS, Bridgend, N/A, United Kingdom
Recall initiated2018-11-01
Classified2019-01-03
Reported by FDA2019-01-09
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PL, PT, RU, SE, SG
Model numbers6801819

Product

VITROS HBeAg Reagent, Cat. No 6801819 Product Usage: For the in vitro qualitative detection of hepatitis B e antigen (HBeAg) in human adult and pediatric (2 to 21 years old) serum from individuals who have symptoms of hepatitis or who may be at risk for hepatitis B virus (HBV) infection using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System. Test results, in conjunction with other serological and clinical information, may be used for the laboratory diagnosis of individuals with acute or chronic hepatitis B, or recovery from hepatitis B infection. The VITROS HBeAg test should not be used to test cord blood samples. Test performance characteristics have not been established in patients under the age of 2, or in populations of immunocompromised or immunosuppressed patients. This test has not been FDA licensed for the screening of blood, plasma and tissue donors.

Reason for recall

The firm is extending their previous July 2018 recall to include additional product codes. Biased results may occur for these additional products at biotin concentrations which are lower than indicated in the current Instructions for Use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0688-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.