Data Foundry

Medical device recalls 2024

Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog… — Class II medical device recall Z-0688-2025

Ongoing recall by CooperSurgical, Inc., reported 2024-12-25, distributed in CA, IL, IN, MT, NC, NJ, NY, PA…. The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperat

Recall numberZ-0688-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2024-11-15
Classified2024-12-18
Reported by FDA2024-12-25
DistributedCA, IL, IN, MT, NC, NJ, NY, PA, SD, TX, VA
Countries outside the USIQ
Quantity37 units
UPC / GTIN / UDI-DI00888937025057
Serial numbers2244KH01, 2244KH02, 2244KH03, 2244KH04, 2244KH05, 2244KH06, 2244KH07, 2244KH08, 2244KH09, 2244KH10, 2313KH01, 2313KH02, 2313KH03, 2313KH04, 2313KH05, 2313KH06, 2313KH07, 2313KH08, 2313KH09, 2313KH10, 2320KH21, 2320KH22, 2320KH23, 2320KH24, 2320KH25 and 10 more
Model numbersK22074

Product

Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: No

Reason for recall

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.