Cobalt Chromium Heads Transcend: 26000017 — Class III medical device recall Z-0689-2017

Terminated recall by MicroPort Orthopedics Inc., reported 2016-12-07, distributed nationwide. Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal

Recall numberZ-0689-2017
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMicroPort Orthopedics Inc., Arlington, TN, United States
Recall initiated2016-06-21
Classified2016-12-01
Reported by FDA2016-12-07
Terminated2017-07-01
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BE, BG, BR, CA, CH, CN, CY, CZ, DE, DK, ES, FI, FR, GB, GR, IR, IT, JM, JP, KR, LT, MY, NL, PH, PL, PT, RU, SI, TR, TW, VN, ZA
Quantity108,362 units
Serial numbers0101000331, 0101001897, 0101010115, 0101010910, 0101010911, 0101010912, 0101010919, 0101012378, 0101013692, 0101013698, 0101013699, 0101013705, 0101016207, 0101016260, 0101016264, 0101016265, 0101016267, 0101016268, 0101017299, 0101017300, 0101017302, 0101018642, 0101018644, 0101018645, 0101018646 and 475 more
Model numbers26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 26000025, 26000026, 26000027, 26000028, 38013800, 38013804, 38013835, 38014000, 38014004, 38014035, 38014200, 38014204, 38014235, 38014400, 38014404, 38014435, 38014600 and 76 more

Product

Cobalt Chromium Heads Transcend: 26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 26000025, 26000026, 26000027, 26000028 Conserve: 38013800, 38013804, 38013835, 38014000, 38014004, 38014035, 38014200, 38014204, 38014235, 38014400, 38014404, 38014435, 38014600, 38014604, 38014635, 38014800, 38014804, 38014835, 38015000, 38015004, 38015035, 38015200, 38015204, 38015235, 38015400, 38015404, 38015435, 38015600, 38015604, 38015635, 38AC3600, 38AC3800, 38AC4000, 38AC4200, 38AC4400, 38AC4600, 38AC4800, 38AC5000, 38AC5200, 38AC5400, 38AC5600, 38AM3600, 38AM3604, 38AM3635, 38AM3800, 38AM3804, 38AM3835, 38AM4000, 38AM4004, 38AM4035, 38AM4200, 38AM4204, 38AM4235, 38AM4400, 38AM4404, 38AM4435, 38AM4600, 38AM4604, 38AM4635, 38AM4800, 38AM4804, 38AM4835, 38AM5000, 38AM5004, 38AM5035, 38AM5200, 38AM5204, 38AM5235, 38AM5400, 38AM5404, 38AM5435, 38AM5600, 38AM5604, 38AM5635, PL38HM4400, PL38HM4404, PL38HM4435, PL38HM4600, PL38HM4604, PL38HM4635, PL38HM4800, PL38HM4804, PL38HM4835, PL38HM5000, PL38HM5004, PL38HM5035, PL38HM5200, PL38HM5204, PL38HM5235

Reason for recall

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.