Data Foundry

Medical device recalls 2024

Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog… — Class II medical device recall Z-0689-2025

Ongoing recall by CooperSurgical, Inc., reported 2024-12-25, distributed in CA, IL, IN, MT, NC, NJ, NY, PA…. The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperat

Recall numberZ-0689-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2024-11-15
Classified2024-12-18
Reported by FDA2024-12-25
DistributedCA, IL, IN, MT, NC, NJ, NY, PA, SD, TX, VA
Countries outside the USIQ
Quantity22 units
UPC / GTIN / UDI-DI00888937025071
Serial numbers2218KH46, 2218KH47, 2218KH48, 2218KH49, 2218KH50, 2218KH51, 2218KH52, 2218KH53, 2218KH54, 2218KH55, 2218KH56, 2218KH57, 2218KH58, 2218KH59, 2218KH60, 2233KH11, 2233KH12, 2233KH13, 2233KH14, 2233KH15, 2233KH16, 2233KH17, 2233KH18, 2233KH19, 2233KH20 and 80 more
Model numbersK23054

Product

Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: No

Reason for recall

The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0689-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.