Data Foundry

Medical device recalls 2022

AU/DxC AU Magnesium — Class II medical device recall Z-0692-2022

Terminated recall by Beckman Coulter Inc., reported 2022-03-09, distributed in AK, AL, AR, AS, AZ, CA, CO, CT…. Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in th

Recall numberZ-0692-2022
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2021-12-08
Classified2022-02-28
Reported by FDA2022-03-09
Terminated2025-01-28
DistributedAK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USAE, AT, AU, BD, BE, BG, BH, BR, CA, CH, CN, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, HR, HU, ID, IE, IL, IN, IT, JO, KE, KR, KW, LB, LK, LT, MA, MX, MY, NG, NL, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, UY, VN, ZA
Quantity181,776 kits
UPC / GTIN / UDI-DI15099590010850
Model numbersOSR6189

Product

AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850

Reason for recall

Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0692-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.