Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation… — Class II medical device recall Z-0694-2018
Terminated recall by Integra LifeSciences Corp., reported 2018-03-07, distributed nationwide. A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when
| Recall number | Z-0694-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Integra LifeSciences Corp., Plainsboro, NJ, United States |
| Recall initiated | 2017-09-26 |
| Classified | 2018-02-23 |
| Reported by FDA | 2018-03-07 |
| Terminated | 2018-08-14 |
| Distributed | Nationwide (US) |
| Countries outside the US | DE, DO, FR, GB, GR, IT, NL, SE, ZA |
| Quantity | 155 |
| Lot numbers | 1885744, 1923802, 1985221, 2057996, 2109587, 2118507 |
| Model numbers | C7000 |
Product
Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses are cleared for use in the US only: Gastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy.
Reason for recall
A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.