A20975A "Working insert — Class II medical device recall Z-0696-2018
Terminated recall by Olympus Corporation of the Americas, reported 2018-03-07, distributed nationwide. Olympus has received complaints about fragments of adhesive which detached from inside the working channel of
| Recall number | Z-0696-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Olympus Corporation of the Americas, Center Valley, PA, United States |
| Recall initiated | 2017-08-18 |
| Classified | 2018-02-23 |
| Reported by FDA | 2018-03-07 |
| Terminated | 2020-07-07 |
| Distributed | Nationwide (US) |
| Quantity | 8,741 |
| Reason classes | packaging, device malfunction |
| Lot numbers | 13-ZW, 13ZW, 146W, 172W, 174W, A20975A, A20976A, A20977A |
Product
A20975A "Working insert, with ramp, one way" (Lots 13ZW to 172W; A20976A "Bridge, one way" (Lots 146W to 172W) and A20977A "Bridge, two way" (Lots 146W to 174W) Product Usage: A20975A: Working insert for endoscopic diagnosis and treatment in urologic applications. A20976A: Bridge for endoscopic diagnosis and treatment in urologic applications. A20977A: Bridge for endoscopic diagnosis and treatment in urologic applications.
Reason for recall
Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping, missing pieces, and delamination of the adhesive have also been reported. In one case, a fragment of the adhesive was detected inside a patient's urethra during a procedure and was retrieved without injury to the patient. The manufacturer, Olympus Winter & lbe GmbH ("OWi") has determined that adhesive can be dislodged during the intended use of the cystoscopy bridge or working insert, e.g., when inserting an instrument through the working channel. As a result, a fragment of the adhesive may fall inside the patient's bladder, urethra or uterus and would need to be retrieved, extending the length of the procedure or requiring additional surgical treatment. The adhesive is used to seal gaps inside the working channel of the cystoscopy bridges and working inserts. In 2013, the adhesive used until then was discontinued and had to be replaced. Current investigations revealed that the durability of the new adhesive is insufficient
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0696-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.