Foot Controls impacted for the following devices: OAKWORKS¿ Medical… — Class II medical device recall Z-0703-2018
Terminated recall by Oakworks Inc, reported 2018-03-07, distributed nationwide. The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 an
| Recall number | Z-0703-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Oakworks Inc, New Freedom, PA, United States |
| Recall initiated | 2017-07-31 |
| Classified | 2018-02-23 |
| Reported by FDA | 2018-03-07 |
| Terminated | 2019-03-29 |
| Distributed | Nationwide (US) |
| Quantity | 1,104 |
| Model numbers | 300, 301, 400, 401, 75205-T01, 75207-T01, 75209-T01, 75212-T01 |
Product
Foot Controls impacted for the following devices: OAKWORKS¿ Medical CFPM 300 Imaging Table OAKWORKS¿ Medical CFPM 301 Imaging Table OAKWORKS¿ Medical CFPM 400 Imaging Table OAKWORKS¿ Medical CFPM 401 Imaging Table OAKWORKS¿ Medical CFPMB 301 Bariatric Imaging Table OAKWORKS¿ Medical CFUR 301 Urology Table OAKWORKS¿ Medical CFUR 401 Urology Table OAKWORKS¿ Medical CFLU 401 Lithotripsy/Urology Table
Reason for recall
The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.