30" — Class II medical device recall Z-0703-2021
Ongoing recall by ICU Medical, Inc., reported 2020-12-30, distributed nationwide. Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may pote
| Recall number | Z-0703-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | ICU Medical, Inc., San Clemente, CA, United States |
| Recall initiated | 2020-08-31 |
| Classified | 2020-12-23 |
| Reported by FDA | 2020-12-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, AU, BE, BR, CA, CH, DE, ES, FI, FR, GB, GR, HK, HR, IE, IL, IT, LU, MY, NL, PL, RO, SE, SG, SK, TH, TR, TW, ZA |
| Quantity | 231,250 |
| Reason classes | packaging |
| Lot numbers | 4560357, 4573189, 4605106, 4606684, 4606697, 4712385, 4714539, 4720364, 4720479, 4724700, 4724701, 4724703, 4724728, 4724743, 4726940, 4733557, 4735662, 4735669, 4735675, 4735696, 4735699, 4738028, 4742091, 4742630, 4742639, 4742641, 4743901, 4749454, 4749474, 4749489, 4752843, 4756268, 4757782, 4758385, 4758414, 4758441, 4763384, 4764898, 4765922, 4765927 and 38 more |
| Model numbers | CL3011, CL3011T, CL3020, CL3028, CL3028T, CL3361, CL3538, CL3929, CL3931, CL3932, CL3935, CL3941, CL3946, CL3948, CL3960, CL3960T, CL3978, CL4100, CL4131, CL4151 |
Product
30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978
Reason for recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0703-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.