QC Sets and Panels: KWIK-STIK Cepheid Smart GBS QC Set — Class II medical device recall Z-0704-2022
Terminated recall by Microbiologics Inc, reported 2022-03-09, distributed nationwide. The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xper
| Recall number | Z-0704-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Microbiologics Inc, Saint Cloud, MN, United States |
| Recall initiated | 2015-12-01 |
| Classified | 2022-03-01 |
| Reported by FDA | 2022-03-09 |
| Terminated | 2022-07-07 |
| Distributed | Nationwide (US) |
| Quantity | 8 units |
| UPC / GTIN / UDI-DI | 70845357028630 |
| Lot numbers | 70845357028630QC, 8165-25 |
| Model numbers | 8165 |
Product
QC Sets and Panels: KWIK-STIK Cepheid Smart GBS QC Set (Live Culture). Sets include: 1. 0864E4 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 2. 0864E5 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 3. 0864E7 Streptococcus agalactiae (Group B, lb) derived from ATCC 12401; 4. Negative Control 0243 Lactobacillus acidophilus derived from ATCC 4356
Reason for recall
The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This causes a failure when used as a quality control for the Cepheid Xpert GBS and Cepheid Smart GBS systems. The failure of the quality control test causes a disruption for the testing laboratory which may result in repeated testing or the need for alternative test methods. This recall is applicable to the 8164 and 8165 sets only.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0704-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.