Multiples Anspach Micro Curved Attachments — Class II medical device recall Z-0708-2014
Terminated recall by The Anspach Effort, Inc., reported 2014-01-15, distributed nationwide. During a recent review of documentation Anspach determined that the current directions for use for the MCA cut
| Recall number | Z-0708-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | The Anspach Effort, Inc., Palm Beach Gardens, FL, United States |
| Recall initiated | 2012-10-26 |
| Classified | 2014-01-09 |
| Reported by FDA | 2014-01-15 |
| Terminated | 2014-06-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CH, ES, FR, GB, IT, JP, NL, ZA |
Product
Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD 1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM. The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting.
Reason for recall
During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0708-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.