King Systems KLTSD414 Quantity — Class II medical device recall Z-0709-2014
Terminated recall by King Systems Corp., reported 2014-01-22, distributed in CA, NY. KLTD214 airways found inside KLTSD414 packages. The KLTSD414 airway has a lumen for use of a gastric suction t
| Recall number | Z-0709-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | King Systems Corp., Noblesville, IN, United States |
| Recall initiated | 2013-12-13 |
| Classified | 2014-01-10 |
| Reported by FDA | 2014-01-22 |
| Terminated | 2014-05-15 |
| Distributed | CA, NY |
| Lot numbers | I15LB |
| Model numbers | KLTSD414 |
Product
King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part Number(s): KLTSD414 KING LTSD KIT/KIT/TROUSSE/KIT SIZE 4 Contents: 1 - KLTSD Size 4......(155 - 180cm) .. 1 - 60cc Syringe 1 - Sterile Lubricant 32-5332 05/10 Intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration.
Reason for recall
KLTD214 airways found inside KLTSD414 packages. The KLTSD414 airway has a lumen for use of a gastric suction tube; the KLTD214 airway does not include a lumen for a gastric suction tube.