Data Foundry

Medical device recalls 2016

Allura Xper FD20/20 — Class II medical device recall Z-0713-2016

Terminated recall by Philips Electronics North America Corporation, reported 2016-02-10, distributed nationwide. The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and s

Recall numberZ-0713-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2015-11-03
Classified2016-01-30
Reported by FDA2016-02-10
Terminated2020-09-15
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MT, MX, MY, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VE, VN, ZA
Quantity13,297 in
Model numbers722038

Product

Allura Xper FD20/20; Model Number: 722038 Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. " Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and ElectroPhysiology " Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Reason for recall

The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0713-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.