Data Foundry

Medical device recalls 2018

SOMATOM Go — Class II medical device recall Z-0713-2018

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2018-03-07, distributed nationwide. The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potenti

Recall numberZ-0713-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2017-10-02
Classified2018-02-23
Reported by FDA2018-03-07
Terminated2018-09-27
DistributedNationwide (US)
Quantity6
Serial numbers108054, 108059, 108062, 108069, 108076, 108077
Model numbers11061628

Product

SOMATOM Go.Up computed tomography (CT) x-ray system table, Model Number: 11061628 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Reason for recall

The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0713-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.