Draeger Infinity Acute Care System Monitoring Solution — Class II medical device recall Z-0714-2014
Terminated recall by Draeger Medical Systems, Inc., reported 2014-01-22, distributed in CA, IL, IN, MA, NH, NY, OK, TX. After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently
| Recall number | Z-0714-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Draeger Medical Systems, Inc., Telford, PA, United States |
| Recall initiated | 2013-12-10 |
| Classified | 2014-01-10 |
| Reported by FDA | 2014-01-22 |
| Terminated | 2015-05-20 |
| Distributed | CA, IL, IN, MA, NH, NY, OK, TX |
| Quantity | 439 |
Product
Draeger Infinity Acute Care System Monitoring Solution
Reason for recall
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.