Data Foundry

Medical device recalls 2014

Draeger Infinity Acute Care System Monitoring Solution — Class II medical device recall Z-0714-2014

Terminated recall by Draeger Medical Systems, Inc., reported 2014-01-22, distributed in CA, IL, IN, MA, NH, NY, OK, TX. After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently

Recall numberZ-0714-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDraeger Medical Systems, Inc., Telford, PA, United States
Recall initiated2013-12-10
Classified2014-01-10
Reported by FDA2014-01-22
Terminated2015-05-20
DistributedCA, IL, IN, MA, NH, NY, OK, TX
Quantity439

Product

Draeger Infinity Acute Care System Monitoring Solution

Reason for recall

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0714-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.