Allura Xper FD and and Xper FD ORT — Class II medical device recall Z-0717-2018
Terminated recall by Philips Electronics North America Corporation, reported 2018-03-07, distributed nationwide. Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage
| Recall number | Z-0717-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Electronics North America Corporation, Andover, MA, United States |
| Recall initiated | 2017-12-20 |
| Classified | 2018-02-24 |
| Reported by FDA | 2018-03-07 |
| Terminated | 2020-06-16 |
| Distributed | Nationwide (US) |
| Quantity | 1,486 |
Product
Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper FD20, Allura Xper FD20/10, Allura Xper FD20/15 and Allura Xper FD10/10 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and ElectroPhysiology. Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room (OR Table)
Reason for recall
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0717-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.