Data Foundry

Medical device recalls 2016

Radiation Therapy Treatment Planning System Product Usage: RayStation… — Class II medical device recall Z-0720-2017

Terminated recall by RAYSEARCH LABORATORIES AB, reported 2016-12-14, distributed nationwide. An error may occur with the display of dose computed on images other than the planning CT (auxiliary CT) when

Recall numberZ-0720-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden
Recall initiated2016-10-11
Classified2016-12-07
Reported by FDA2016-12-14
Terminated2019-06-11
DistributedNationwide (US)
Quantity540 units

Product

Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

Reason for recall

An error may occur with the display of dose computed on images other than the planning CT (auxiliary CT) when using multiple patient cases in RayStation 5. If a CT image set with the same Frame of Reference as the displayed auxiliary CT set exists in another case, the dose display may be incorrect. When the error occurs, the display of dose in patient views, including the maximum dose position, the dynamic isodose lines and the dose grid may be incorrect. The dose value normally displayed in the upper left corner when pointing in the 2D view may be incorrect or missing.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0720-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.