Versys 6 inch Beaded Full Coat Plus Hip Prosthesis — Class II medical device recall Z-0722-2018
Terminated recall by Zimmer Biomet, Inc., reported 2018-03-07, distributed nationwide. The products being recalled exceeded the weight previously tested leading to a possibility of damage to the pa
| Recall number | Z-0722-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Zimmer Biomet, Inc., Warsaw, IN, United States |
| Recall initiated | 2017-08-01 |
| Classified | 2018-02-26 |
| Reported by FDA | 2018-03-07 |
| Terminated | 2020-03-18 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, AU, BR, BS, CA, CH, CN, DE, DO, ES, FR, GB, IL, JP, KR, LB, MX, MY, NL, PA, SG, TH |
| Quantity | 38,069 units |
| Reason classes | sterility |
| Model numbers | 00784301406, 00784301506, 00784301606, 00784301706, 00784301746, 00784301756, 00784301806, 00784301826, 00784301836, 00784301846, 00784301856, 00784301906, 00784301926, 00784301936, 00784301946, 00784301956, 00784302006, 00784302026, 00784302036, 00784302046, 00784302056, 00784302106, 00784302126, 00784302136, 00784302146 and 7 more |
Product
Versys 6 inch Beaded Full Coat Plus Hip Prosthesis. Labeled as follows: a. VERSYS 6 INCH BEADED FC 19X160MM STD BODY STD NECK (Item No. 00894301906); b. VERSYS 6 INCH BEADED FC 21X160MM STD BODY STD NECK (Item No. 00784302106); c. VERSYS 6 INCH BEADED FC 22X160MM STD BODY STD NECK (Item No. 00784302206); d. VERSYS 6 INCH BEADED FC STEM 19X160MM LM (Item No. 00784301936); e. VERSYS 6 INCH BEADED FC STEM 20X160MM LM (Item No. 00784302036); f. VERSYS 6 INCH BEADED FC STEM 21X160MM LM (Item No. 00784302136); g. VERSYS 6 INCH BEADED FC STEM 22X160MM LM (Item No. 00784302236); h. VERSYS 6 INCH BEADED FC 14X160MM STD BODY STD NECK (Item No. 00784301406); i. VERSYS 6 INCH BEADED FC 15X160MM STD BODY STD NECK (Item No. 00784301506); j. VERSYS 6 INCH BEADED FC 16X160MM STD BODY STD NECK (Item No. 00784301606); k. VERSYS 6 INCH BEADED FC 17X160MM STD BODY STD NECK (Item No. 00784301706); l. VERSYS 6 INCH BEADED FC 18X160MM STD BODY STD NECK (Item No. 00784301806); m. VERSYS 6 INCH BEADED FC 18X160MM STD BODY EXT NECK (Item No. 00784301826); n. VERSYS 6 INCH BEADED FC 19X160MM STD BODY EXT NECK (Item No. 00784301926); o. VERSYS 6 INCH BEADED FC 20X160MM STD BODY STD NECK (Item No. 00784302006); p. VERSYS 6 INCH BEADED FC 20X160MM STD BODY EXT NECK (Item No. 00784302026); q. VERSYS 6 INCH BEADED FC 21X160MM STD BODY EXT NECK (Item No. 00784302126); r. VERSYS 6 INCH BEADED FC 22X160MM STD BODY EXT NECK (Item No. 00784302226); s. VERSYS 6 INCH BEADED FC 17X160MM LM BODY XEXT NECK (Item No. 00784301746); t. VERSYS 6 INCH BEADED FC 17X160MM LM BODY EXT NECK (Item No. 00784301756 u. VERSYS 6 INCH BEADED FC STEM 18X160MM LM (Item No. 00784301836); v. VERSYS 6 INCH BEADED FC 18X160MM LM BODY XEXT NECK (Item No. 00784301846); w. VERSYS 6 INCH BEADED FC 18X160MM LM BODY EXT NECK (Item No. 00784301856); x. VERSYS 6 INCH BEADED FC 19X160MM LM BODY XEXT NECK (Item No. 00784301946); y. VERSYS 6 INCH BEADED FC 19X160MM LM BODY EXT NECK (Item No. 00784301956); z. VERSYS 6 INCH BEADED FC 20X160MM LM BODY XEXT NECK (Item No. 00784302046); aa. VERSYS 6 INCH BEADED FC 20X160MM LM BODY EXT NECK (Item No. 00784302056); ab. VERSYS 6 INCH BEADED FC 21X160MM LM BODY XEXT NECK (Item No. 00784302146); ac. VERSYS 6 INCH BEADED FC 21X160MM LM BODY EXT NECK (Item No. 00784302156); ad. VERSYS 6 INCH BEADED FC 22X160MM LM BODY XEXT NECK (Item No. 00784302246); ae. VERSYS 6 INCH BEADED FC 22X160MM LM BODY EXT NECK (Item No. 00784302256) Components of the system are intended to replace the proximal femur, mid-shaft femur, distal femur, proximal tibia, and/or total knee in cases that require extensive resection and restoration.
Reason for recall
The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0722-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.