Philips Allura Xper — Class II medical device recall Z-0724-2022
Ongoing recall by Philips North America Llc, reported 2022-03-16, distributed nationwide. Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If
| Recall number | Z-0724-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips North America Llc, Cambridge, MA, United States |
| Recall initiated | 2021-11-24 |
| Classified | 2022-03-04 |
| Reported by FDA | 2022-03-16 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, BD, BR, CH, CL, CN, DE, DK, DO, ES, FR, GB, GR, ID, IL, IN, IR, IT, JP, KR, LK, MX, MY, NL, PH, SA, TH, TR, TW, ZA |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 00884838054189, 00884838054202, 00884838054219 |
| Model numbers | 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039 |
Product
Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Reason for recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.