Data Foundry

Medical device recalls 2022

Dispensing Pins for Air Inlet Filter — Class II medical device recall Z-0727-2022

Terminated recall by B. Braun Medical, Inc., reported 2022-03-16, distributed nationwide. This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inle

Recall numberZ-0727-2022
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmB. Braun Medical, Inc., Allentown, PA, United States
Recall initiated2022-01-31
Classified2022-03-07
Reported by FDA2022-03-16
Terminated2024-04-29
DistributedNationwide (US)
Countries outside the USCA
Reason classespackaging
UPC / GTIN / UDI-DI04022495769633
Lot numbers0061788969
Model numbers412005, DP2500
Expiration dates2026-05-31

Product

Dispensing Pins for Air Inlet Filter, Product Catalog No. 412005, DP2500 Dispensing Pin W/Valve. A valve for aspiration or injection of medication or enteral feeding solution from inverted multi-dose vials.

Reason for recall

This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the dose of medication delivered to the patient, cause delays in treatment due to the need to replace the device or may cause contamination of the fluid path that can lead to bloodstream infection.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0727-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.