Data Foundry

Medical device recalls 2020

S — Class II medical device recall Z-0728-2020

Ongoing recall by Cardinal Health Inc., reported 2020-01-01, distributed nationwide. Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information

Recall numberZ-0728-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCardinal Health Inc., Dublin, OH, United States
Recall initiated2019-11-14
Classified2019-12-20
Reported by FDA2020-01-01
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GT, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LT, LU, LV, MA, MT, MX, MY, NL, NO, NZ, OM, PA, PE, PH, PL, PT, QA, SE, SK, TH, TR, UA, UY, VE, ZA
Reason classeslabeling

Product

S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) or 7 French (2.3 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The SMART Control stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy) material and a 6 French (2.0 mm) sheathed delivery system. The stent is laser cut from a solid nitinol tube into a fine mesh ( Z Configuration) design. The stent expands to its unconstrained diameter when released from the deployment catheter into the iliac artery. Upon deployment, the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2 mm smaller in diameter than the unconstrained diameter of the stent. The stent is equipped with 12 tantalum markers, 6 at each end, to increase the visibility of the stent under fluoroscopic imaging. The SMART Control stent system is designed to be delivered through an appropriately sized catheter sheath introducer.

Reason for recall

Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0728-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.