Data Foundry

Medical device recalls 2022

YSIO X — Class II medical device recall Z-0730-2022

Ongoing recall by Siemens Medical Solutions USA, Inc, reported 2022-04-13, distributed nationwide. For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for

Recall numberZ-0730-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedFDA mandated
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2022-03-07
Classified2022-04-05
Reported by FDA2022-04-13
DistributedNationwide (US)
Quantity53 units

Product

YSIO X.Pree

Reason for recall

For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0730-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.