Lumex — Class II medical device recall Z-0732-2017
Terminated recall by GF Health Products, Inc, reported 2016-12-21, distributed nationwide. GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecti
| Recall number | Z-0732-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GF Health Products, Inc, Fond Du Lac, WI, United States |
| Recall initiated | 2016-09-30 |
| Classified | 2016-12-12 |
| Reported by FDA | 2016-12-21 |
| Terminated | 2019-04-29 |
| Distributed | Nationwide (US) |
| Countries outside the US | BS, CA, CL, SV |
| Reason classes | packaging |
Product
Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912
Reason for recall
GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.