Data Foundry

Medical device recalls 2016

Lumex — Class II medical device recall Z-0732-2017

Terminated recall by GF Health Products, Inc, reported 2016-12-21, distributed nationwide. GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecti

Recall numberZ-0732-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGF Health Products, Inc, Fond Du Lac, WI, United States
Recall initiated2016-09-30
Classified2016-12-12
Reported by FDA2016-12-21
Terminated2019-04-29
DistributedNationwide (US)
Countries outside the USBS, CA, CL, SV
Reason classespackaging

Product

Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912

Reason for recall

GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.