Data Foundry

Medical device recalls 2014

MCA5-06SD — Class II medical device recall Z-0751-2014

Terminated recall by The Anspach Effort, Inc., reported 2014-01-22, distributed nationwide. The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and

Recall numberZ-0751-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmThe Anspach Effort, Inc., Palm Beach Gardens, FL, United States
Recall initiated2012-03-29
Classified2014-01-14
Reported by FDA2014-01-22
Terminated2015-03-11
DistributedNationwide (US)
Quantity57
Reason classeslabeling
Lot numbers0.6MM, MCA5-06SD

Product

MCA5-06SD - 0.6MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason for recall

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide