Data Foundry

Medical device recalls 2025

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer… — Class II medical device recall Z-0753-2025

Ongoing recall by Abbott Medical, reported 2025-01-01, distributed nationwide. Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence

Recall numberZ-0753-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbbott Medical, Sylmar, CA, United States
Recall initiated2024-11-21
Classified2024-12-23
Reported by FDA2025-01-01
DistributedNationwide (US)
Countries outside the USAE, AT, BE, CA, CH, DE, DK, ES, FI, FR, GB, HK, IL, IT, KW, NL, NO, PL, PT, SA, SE
Quantity7,295 programmers
Reason classessoftware
UPC / GTIN / UDI-DI05414734509725
Serial numbers10764, 11353, 11364, 11371, 11373, 11387, 11513, 11522, 11577, 11605, 11628, 11635, 11637, 11646, 11653, 11682, 11688, 11700, 11708, 11743, 11746, 11756, 11760, 11785, 11790 and 475 more
Model numbers3330, 3650, MER3700

Product

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Reason for recall

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0753-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.