Data Foundry

Medical device recalls 2016

ADVIA Chemistry XPT Systems — Class II medical device recall Z-0755-2016

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2016-02-10, distributed nationwide. Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control

Recall numberZ-0755-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States
Recall initiated2015-10-07
Classified2016-02-04
Reported by FDA2016-02-10
Terminated2018-05-22
DistributedNationwide (US)
Countries outside the USAU, BH, DE, DK, EG, ES, FI, FR, GB, IT, KR, NL, NO, NZ, SG, SK
Quantity87 units
Reason classesdevice malfunction, software

Product

ADVIA Chemistry XPT Systems, Software Version 1.0.3, SMN 11127538 Clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.

Reason for recall

Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS) communication/Laboratory Automation (LAS) issue, Printer Driver Resets, ISE Calibration Ranges are too conservative for Urine Sodium, Archiving and deletion may fail and Workstation services may restart.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0755-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.