Cardinal Health Suction Canister Liner — Class II medical device recall Z-0756-2024
Ongoing recall by Cardinal Health 200, LLC, reported 2024-01-24. The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of sucti
| Recall number | Z-0756-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cardinal Health 200, LLC, Waukegan, IL, United States |
| Recall initiated | 2023-12-21 |
| Classified | 2024-01-18 |
| Reported by FDA | 2024-01-24 |
| Countries outside the US | AU, CA, NZ |
| Quantity | 74,515 |
| UPC / GTIN / UDI-DI | 50630140152670 |
| Lot numbers | J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618 |
| Model numbers | 65651-517 |
Product
Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517
Reason for recall
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.