The LIFEPAK 15 monitor/defibrillator is intended for use by trained… — Class II medical device recall Z-0761-2017
Terminated recall by Physio-Control, Inc., reported 2016-12-21. Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedl
| Recall number | Z-0761-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Physio-Control, Inc., Redmond, WA, United States |
| Recall initiated | 2010-03-11 |
| Classified | 2016-12-14 |
| Reported by FDA | 2016-12-21 |
| Terminated | 2016-12-16 |
| Model numbers | V-15-2-000XXX |
Product
The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX
Reason for recall
Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.