Edwards Lifesciences Swan Ganz Thermodilution Catheter — Class I medical device recall Z-0765-2019
Terminated recall by Edwards Lifesciences, LLC, reported 2019-02-06. These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens ar
| Recall number | Z-0765-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Edwards Lifesciences, LLC, Irvine, CA, United States |
| Recall initiated | 2018-12-21 |
| Classified | 2019-01-31 |
| Reported by FDA | 2019-02-06 |
| Terminated | 2023-11-02 |
| Quantity | 459 units |
| Lot numbers | 61176367, 61176370 |
| Model numbers | 151F7 |
Product
Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Reason for recall
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.