Data Foundry

Medical device recalls 2019

Edwards Lifesciences Swan Ganz Thermodilution Catheter — Class I medical device recall Z-0765-2019

Terminated recall by Edwards Lifesciences, LLC, reported 2019-02-06. These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens ar

Recall numberZ-0765-2019
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmEdwards Lifesciences, LLC, Irvine, CA, United States
Recall initiated2018-12-21
Classified2019-01-31
Reported by FDA2019-02-06
Terminated2023-11-02
Quantity459 units
Lot numbers61176367, 61176370
Model numbers151F7

Product

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Reason for recall

These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0765-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.