Data Foundry

Medical device recalls 2020

The PLEX Elite 9000 — Class II medical device recall Z-0770-2020

Terminated recall by Carl Zeiss Meditec, Inc., reported 2020-01-08, distributed nationwide. Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contac

Recall numberZ-0770-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCarl Zeiss Meditec, Inc., Dublin, CA, United States
Recall initiated2018-10-16
Classified2020-01-02
Reported by FDA2020-01-08
Terminated2021-06-11
DistributedNationwide (US)
Countries outside the USAE, AT, AU, CA, CH, CN, DE, FR, IL, IN, IT, JP, KR, NO, SG, TW
Quantity151 devices
Serial numbersPE9000-001, PE9000-0010, PE9000-0011, PE9000-0012, PE9000-0013, PE9000-0014, PE9000-0015, PE9000-0016, PE9000-002, PE9000-0020, PE9000-0021, PE9000-0022, PE9000-0023, PE9000-0024, PE9000-0025, PE9000-0026, PE9000-0027, PE9000-0029, PE9000-003, PE9000-0030, PE9000-0031, PE9000-0032, PE9000-0033, PE9000-0034, PE9000-0035 and 126 more

Product

The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.

Reason for recall

Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.