Data Foundry

Medical device recalls 2014

Arena-L Trial Handle — Class II medical device recall Z-0776-2014

Terminated recall by SpineFrontier, Inc., reported 2014-01-22. Handles and heads may not be compatible.

Recall numberZ-0776-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSpineFrontier, Inc., Beverly, MA, United States
Recall initiated2013-09-24
Classified2014-01-16
Reported by FDA2014-01-22
Terminated2016-08-04
Quantity23 devices
Model numbers11-81002

Product

Arena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of the implant, manufactured from 17-4 Stainless Steel into 35X26mm (11-81004-XX) and 38X28mm (11-81012-XX) varieties, each with multiple height configurations. The Trial Heads are intended to attach tot he Trial Handle (11-81002) and to be inserted into the intervertebral disc space. The Trial Handle consists of a knurled handle and cylindrical shaft with an M4 threads interface at the distal end; 17-4 Stainless steel is used as the material for the handles. The Trial Heads distract the disc space to the surgeon's desired height to gauge which size implant should be utilized to complete the Anterior Lumbar Interbody Fusion Procedure.

Reason for recall

Handles and heads may not be compatible.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0776-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.