Arena-L 38x28Trial Head I 10¿ Lordotic — Class II medical device recall Z-0778-2014
Terminated recall by SpineFrontier, Inc., reported 2014-01-22. Handles and heads may not be compatible.
| Recall number | Z-0778-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | SpineFrontier, Inc., Beverly, MA, United States |
| Recall initiated | 2013-09-24 |
| Classified | 2014-01-16 |
| Reported by FDA | 2014-01-22 |
| Terminated | 2016-08-04 |
| Quantity | 23 devices |
| Model numbers | 11-81012-08, 11-81012-10, 11-81012-12, 11-81012-14 |
Product
Arena-L 38x28Trial Head I 10¿ Lordotic; Part: 11-81012-08, 11-81012-10, 11-81012-12, 11-81012-14 The trial heads are smooth plates that mimic the footprint of the implant, manufactured from 17-4 Stainless Steel into 35X26mm (11-81004-XX) and 38X28mm (11-81012-XX) varieties, each with multiple height configurations. The Trial Heads are intended to attach tot he Trial Handle (11-81002) and to be inserted into the intervertebral disc space. The Trial Handle consists of a knurled handle and cylindrical shaft with an M4 threads interface at the distal end; 17-4 Stainless steel is used as the material for the handles. The Trial Heads distract the disc space to the surgeon's desired height to gauge which size implant should be utilized to complete the Anterior Lumbar Interbody Fusion Procedure
Reason for recall
Handles and heads may not be compatible.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0778-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.