Trilogy 100 — Class I medical device recall Z-0785-2022

Ongoing recall by Philips Respironics, Inc., reported 2022-03-30, distributed nationwide. In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices a

Recall numberZ-0785-2022
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips Respironics, Inc., Murrysville, PA, United States
Recall initiated2018-06-12
Classified2022-03-18
Reported by FDA2022-03-30
DistributedNationwide (US)
Quantity229,353 devices

Product

Trilogy 100, Trilogy 200, Garbin Plus, Aeris, LifeVent Continuous Ventilators The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may be used for both invasive and non-invasive ventilation. It is not intended to be used as a transport ventilator.

Reason for recall

In 2018, Philips Respironics added foam replacement to the preventive maintenance schedule for these devices after discovering the polyester-based polyurethane (PE-PUR) foam may degrade into particles. This 2018 correction has been superseded by the device removal and rework described in Philips Respironics 2021 recall action for the same devices.