Medtronic SynchroMed II Implantable Drug Infusion System — Class I medical device recall Z-0788-2017
Terminated recall by Medtronic Neuromodulation, reported 2017-03-08, distributed nationwide. Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed
| Recall number | Z-0788-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2016-10-03 |
| Classified | 2017-03-01 |
| Reported by FDA | 2017-03-08 |
| Terminated | 2022-06-02 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, CH, DE, FR, GB, GR, IT, MT, NO |
| Quantity | 22,298 software |
| Reason classes | labeling, software |
| Model numbers | 8637, 8840, 8870 |
Product
Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 SynchroMed II Implantable Drug Infusion Pump (supplied in 20 ml or 40 ml reservoir size.) Product Usage: The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids: Chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain, Chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of severe chronic pain, Chronic intrathecal infusion of Lioresal Intrathecal (baclofen Injection ) in the management of serve spasticity, Chronic Intravascular Infusion of Floxuridine (FUDR) or methotrexate for the treatment of primary or metastatic cancer
Reason for recall
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address the priming bolus issue.