Data Foundry

Medical device recalls 2019

Alinity i Processing Module — Class II medical device recall Z-0789-2019

Terminated recall by Abbott Gmbh & Co. KG, reported 2019-02-06. Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in

Recall numberZ-0789-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAbbott Gmbh & Co. KG, Wiesbaden, N/A, Germany
Recall initiated2018-12-20
Classified2019-01-30
Reported by FDA2019-02-06
Terminated2022-09-19
Quantity434 units
Serial numbersAI01001, AI01455, AI01457, AI01459
Model numbers03R65-01, A-30108552-01, A-30109694-01

Product

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Reason for recall

Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0789-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.