CoaguChek XS Prof — Class II medical device recall Z-0803-2021
Terminated recall by Roche Diagnostics Operations, Inc., reported 2021-01-13, distributed nationwide. Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions
| Recall number | Z-0803-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Operations, Inc., Indianapolis, IN, United States |
| Recall initiated | 2020-12-02 |
| Classified | 2021-01-07 |
| Reported by FDA | 2021-01-13 |
| Terminated | 2022-06-30 |
| Distributed | Nationwide (US) |
| Quantity | 212,093 units |
| Model numbers | 04837975001, 06298176001, 08468699001, 08468745001, 08468753001 |
Product
CoaguChek XS Prof (Professional) Meters- IVD professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. Catalog: 04837975001 Units could consist of any of the following strip configurations: 6 count, 24 count, 2x 24 count (i.e. 48 strips)* *CoaguChek XS 3x48 promo LTC HH Catalog: 06298176001 CoaguChek XS 6x24 promo LTC HH Catalog:08468699001 CoaguChek XS 4x48 promo Catalog: 08468745001 CoaguChek XS 6x48 promo Catalog: 08468753001
Reason for recall
Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0803-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.