GE Healthcare — Class II medical device recall Z-0808-2014

Terminated recall by GE Healthcare, LLC, reported 2014-01-29, distributed nationwide. Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on

Recall numberZ-0808-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2013-12-05
Classified2014-01-22
Reported by FDA2014-01-29
Terminated2016-07-25
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BH, BM, BR, CA, CH, CL, CN, CO, CR, CZ, DE, EG, GB, HR, IE, IL, IQ, JM, JO, JP, KR, KW, LB, LT, NG, NZ, PA, PE, PK, PL, PY, QA, RU, SE, SK, TN, VE, ZA
Serial numbersA0HB3179GR, A0HB3216GR, A0HD3119F, A1HB7183G, A1HB7547G, A1HB7598G, A1HB7601G, A1HB7604G, A2TW1861G, A3T15433G, A4S20471G, A4S20482G, A4T18843G, A4T18844G, A4T18846G, A4T18847G, A4T18848G, A4T18850G, A4T18856G, A4T18857G, A4T18858G, A4T18859G, A4T18860G, A4T18861G, A4T18865G and 475 more

Product

GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. The TRAM module functions as a transport monitor during transport, and in the bedside monitor during ICU or OR monitoring. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. TRAM modules acquire, process and store information regarding these parameters. TRAM modules incorporate different monitoring capabilities based on their configuration. K011000: The TRAM 2001 Module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The TRAM 2001 Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric, and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM 2001 Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM 2001 Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport. K053121: The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.

Reason for recall

Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.