Actuator for angiographic x-ray monitor ceiling suspension system — Class II medical device recall Z-0810-2016

Terminated recall by Philips Electronics North America Corporation, reported 2016-02-24, distributed nationwide. The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest

Recall numberZ-0810-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2015-07-06
Classified2016-02-16
Reported by FDA2016-02-24
Terminated2017-02-03
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IQ, IS, IT, JO, JP, KR, KY, LB, LK, LT, MA, MX, MY, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, RO, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, VN, ZA
Quantity3,158 devices
Reason classesdevice malfunction

Product

Actuator for angiographic x-ray monitor ceiling suspension system. Actuator short Assy: THC8AWDS-121 and Actuator long Assy: THC8AWDS-122 used with the following products: 722001 Allura Xper FD 10 C; 722003 Allura Xper FD10; 722005 Allura Xper FD10/10; 722006 Allura Xper FD20; 722008 Allura Xper FD20 Biplane; 722010 Allura Xper FD10; 722011 Allura Xper FD10/10; 722012 Allura Xper FD20; 722013 Allura XPER FD20 BIPLANE; 722015 Allura Xper FD20 OR Table; 722020 Allura Xper FD20 Biplane OR Table; 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table; 722025 Allura Xper FD20 Biplane OR Table; 722026 Allura Xper FD10; 722027 Allura Xper FD10/10; 722028 Allura Xper FD20; 722029 Allura Xper FD20/10; 722035 Allura Xper FD20 OR Table; 722038 Allura Xper FD20/20; 722039 Allura Xper FD20/20 OR Table; 722058 Allura Xper FD20/15; 722123 Field extensions Xper cardio R7.6; 722124 Field extensions Xper vascular R7.6; 722133 Field ext. Xper cardio systems R7.6; 722134 Field ext. Xper vascular systems R7.6; 72249 INTEGRIS BV5000; 722043 INTEGRIS Allura 15-12 (mono); 722121 Field extensions INTEGRIS cardio systems

Reason for recall

The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.