GlideScope Video Laryngoscope Reusable Blades — Class II medical device recall Z-0811-2016

Terminated recall by Verathon, Inc., reported 2016-02-24. Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitatio

Recall numberZ-0811-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmVerathon, Inc., Bothell, WA, United States
Recall initiated2015-04-13
Classified2016-02-17
Reported by FDA2016-02-24
Terminated2016-10-26
Reason classeschemical contamination
Model numbers0574-0007, 0574-0118, AC131852, AC151510, MD131859, MD151622

Product

GlideScope Video Laryngoscope Reusable Blades (GVL 3, 4, 5 and AVL 2, 3, 4, 5). GVL: The GlideScope Video Laryngoscope is a video laryngoscope that incorporates a high-resolution color camera, an LED light source, a rechargeable lithium battery, and video output to a remote monitor. It consists of a video laryngoscope and a portable video monitor. AVL: The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.

Reason for recall

Verathon has revised the Operations and Maintenance Manual (OMM) to specifically state compatibility limitations of the reusable video laryngoscope Blade validated through simulated use cycles, such that exceeding this number of cycles in the specific chemical agent may affect the potential life of the product.