GE Healthcare Optima XR240amx — Class II medical device recall Z-0813-2025
Ongoing recall by GE Medical Systems, LLC, reported 2025-01-15. GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacem
| Recall number | Z-0813-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Medical Systems, LLC, Waukesha, WI, United States |
| Recall initiated | 2024-11-25 |
| Classified | 2025-01-07 |
| Reported by FDA | 2025-01-15 |
| Quantity | 120 units |
| UPC / GTIN / UDI-DI | 00840682122351 |
| Serial numbers | 00130MOB16, 00158MOB17, 082407100689, 082407250223, 206223OPTIMA1, 206248OPT2, 206341OPT2402, 210METM2, 212838OPTIMA1, 219326AMX2, 228867XR240, 228867XR240C, 252847PP241, 256543OPT, 256543OPT1, 256HHPORT5, 256HHPORT6, 260824XR240, 270793OPT, 270793OPT1, 281274OPT1, 281338CL240, 281VHVXR240, 304256OPTIMA240, 305682DIG1 and 94 more |
Product
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Reason for recall
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.