Data Foundry

Medical device recalls 2014

MRI Low Profile Single Lumen port with Open-Ended 6 — Class II medical device recall Z-0814-2014

Terminated recall by Bard Access Systems, reported 2014-01-29, distributed nationwide. Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a

Recall numberZ-0814-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBard Access Systems, Salt Lake City, UT, United States
Recall initiated2013-12-06
Classified2014-01-23
Reported by FDA2014-01-29
Terminated2014-08-05
DistributedNationwide (US)
Countries outside the USBE, CA
Quantity1,358
Reason classeslabeling
Lot numbersREXC0736, REXC0931

Product

MRI Low Profile Single Lumen port with Open-Ended 6.6Fr Silicone Catheter. Product Code 0603880. Product Usage: The 6.6 Fr Open-Ended Single-Lumen venous catheters is packaged into various Bard Acess Systems finished goods as Attachable or Preconnected catheters in BardPort Implanted Port Kits These products are used for patient therapies requiring repeated access to the vascular system. They are indicated for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. Subject ports are not indicated to Power Injection. Port access is performed by percutaneous needle insertion using a non-coring needle. The system consists of two primary components: an injection port with a self-sealing silicone septum and a radiopaque silicone catheter.

Reason for recall

Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0814-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.