Data Foundry

Medical device recalls 2014

EPIQ 7 Ultrasound System — Class II medical device recall Z-0817-2015

Terminated recall by Philips Ultrasound, Inc., reported 2014-12-24, distributed in AK, CA, CO, CT, DC, DE, FL, GA…. When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications,

Recall numberZ-0817-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Ultrasound, Inc., Bothell, WA, United States
Recall initiated2014-11-05
Classified2014-12-16
Reported by FDA2014-12-24
Terminated2016-08-15
DistributedAK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, WI
Countries outside the USAE, AR, AT, AU, BE, CA, CH, CO, DE, EC, EE, ES, FI, FR, GB, GR, HK, IL, IN, IT, JP, LU, MX, MY, NL, NO, NZ, PE, PH, SA, SE, SG, SI, TH, TR, TW, VN
Serial numbers079500318
Model numbers795200, 795201, 989605386721

Product

EPIQ 7 Ultrasound System, EPIQ 7 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 7G, EPIC 7C, EPIQ 7W; Catalog Number: 795200 / 795201 and Part Number: 989605386721. Diagnostic Ultrasound Imaging.

Reason for recall

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0817-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.